Contents

Search


flatline protocol

Electrocardiogram registers a flat line. Differential diagnosis: 1) batteries dead 2) monitor gain to low 3) monitor cable not connected to patient, 3-lead connector, or monitor 4) ventricular fibrillation with a sustained vector perpendicular to the monitor Procedure: 1) check power to both defibrillator & monitor 2) check all connections 3) check gain or sensitivity 4) check setting for 'LEAD SELECT' - check each of leads 5) if using 'quick-look' paddles, move paddles 90 degrees a) move paddle at right upper sternal border to left upper sternal border b) move paddle at left apex to right lower sternal border 6) if using adhesive pads a) do NOT unstick them & rotate b) connect 3 limb leads

Related

advanced cardiac life support (ACLS)

General

protocol

References

ACLS - The Reference Texbook ACLS: Principles & Practice, Cummins RO et al (eds), American Heart Association, 2003 ISBN 0-87493-341-2